Estimation of Ferrous Ammonium Sulfate (FAS) using standard Potassium Dichromate solution Potentiometrically.


Apparatus and Materials Required:
  1. Potentiometer
  2. Platinum electrode (indicator electrode)
  3. Calomel or Silver/Silver Chloride electrode (reference electrode)
  4. Burette (50 mL)
  5. Pipette (10 mL)
  6. Volumetric flask (100 mL)
  7. Beaker (250 mL)
  8. Magnetic stirrer with stir bar
  9. Wash bottle
  10. Funnel
  11. Glass rod

Introduction

Using a suitable electrode that reacts to the concentration change during the titration, one can quantify the change in potential in a potentiometric titration. An indicator electrode is an electrode that reacts to a change in the amount of ions present in the solution. The reference electrode, whose potential remains constant during the titration, is paired with the indicator electrode to create a cell. The cell's e.m.f. varies gradually up until the end point, then quickly right before it and again gradually after. The first derivative curve is plotted against V, the peak of which is (∆E/∆V).

The titration's end point is indicated by the curve.

Cr6+ + 3Fe2+ → 3Fe3+ + Cr3+ [Acidic Medium]

The potentiometric titration of FAS (Mohr’s salt) solution with K2Cr2O7 in the presence of H2SO4 is a redox titration. An oxidation-reduction electrode forms when the same material is present in both its reduced and oxidized forms in a solution. This electrode potential can be detected by dipping a platinum wire into the solution. Because of this, the solution contains both Fe2+ and Fe2+ ions when the titration is started, creating an electrode potential that a Pt wire detects. The indicator electrode, then, is (Pt/Fe3+, Fe2+), and its electrode potential is provided by,

E=E0 + 2.303 RT log (Fe3+)/ nF (Fe2+)